CLEARING MALFUNCTIONS
7
Probe
EL. too hot
M
ainte-
nance
Electronics too hot. Ambient temperature too
high?
Allow device to cool down.
Air purge low
Volume flow below set limit.
Check purge air supply.
Filter watch
Flow monitor.
Check purge air supply.
p no signal
No signal from pressure sensor.
Check purge air supply.
p out of range
Sample gas pressure < 500 or > 1200 hPa
(mbar).
---
t air no signal
Broken sensor.
Please contact SICK Customer Service.
[t] no signal
Broken sensor.
EEPROM defect
EEPROM defective.
Heat no signal
Heater fault.
Heater < 1.5 A
Heater defect
Heating too low
No com.
Communication error to optic head and/or
reflector.
Check connection cables.
System
Systemstart
X
tended This message is entered during each system
start.
Shows when last system start was made.
Zero adjust
The start of an adjustment is recorded in the
logbook.
Shows when last adjustment was made.
Boxmeasuring
The start of filter box measurement is recorded
in the logbook.
Shows when last filter box measurement was made.
Reflector search
Search for reflector failed
Check alignment,
matic optical alignment check; option)”, page
Reflector contaminated or defective.
Light intensity weakness too strong on measuring
path.
P
Substitute value
M
ainte-
nance
Calculation is made with a substitute value
because of a pressure measurement error.
The set input (probe, analog input, SCU) shows errors
and the substitute value is therefore used for
calculation.
T
Substitute value
M
ainte-
nance
Calculation is made with a substitute value
because of a temperature measurement error.
The set input (probe, analog input, SCU) of pressure
measurement shows errors and the substitute value
is therefore used for calculation.
Initia-
tor
[1]
Text
Classifi-
cation
Description
Possible causes/remedy
[2]
Table 18
Error messages