- 33 -
Press the START/STOP key in the FADJ screen. The flow rate
conversion temperature is displayed flashing. The value can be switched
between 20.0°C and 25.0°C by pressing the UP/DOWN key. Pressing
the MODE key in this screen cancels calibration and returns to the FADJ
screen.
After setting the flow rate conversion temperature, press the
START/STOP key. The flow volume is displayed flashing. Set the flow
rate to be calibrated by pressing the UP/DOWN key.
Pressing the START/STOP key offsets
the flow rate to zero and starts pump
operation. The operation display LED
flashes (red), the pump operation icon
flashes, and no other operations are accepted for one minute. When the operation display LED
starts flashing (green) and the pump operation icon stops flashing and lights, perform
measurement using a standard flow meter.
When measurement is completed, press the START/STOP key to stop
pump operation.
Match the flow rate value to the number indicated on the standard flow
meter by pressing the UP/DOWN key.
When the above calibration procedure is finished, press the
START/STOP key. "END" is displayed and the screen returns to the
FADJ screen.
* There is no need to calibrate the flow rate each time that the pump is
used.
* Do not turn the pump OFF during calibration. Doing so might cause
malfunction.