- 32 -
Backlight ON Time
The time up to when the backlight turns OFF out after a button on the
control panel is pressed can be set. (See page 29 "Backlight ON
Setting.") The default is 30 seconds.
Holding down the MODE key in this screen causes the SET segment
of the mode display bar to light and the value to flash. In this state, the
value can be set. Change the value by using the UP/DOWN key. After
changing the value, press the START/STOP key. The display value
lights, and the SET segment of the mode display bar goes out to
indicate that setup is complete.
Calibration
Calibration with suction holder connected to suction port in order to do highly precise measurement.
Using Sibata's flow calibrator FC-M1 as a standard flow meter allows easier correction. Refer to the
instruction of flow calibrator FC-M1 for more detail.
FC-M1 and USB cable allows the automatic operation as below:
Pressing the UP/DOWN key after entering the CAL screen enters the flow volume calibration mode.
* Calibration here is 1-point calibration. So, accuracy will deviate with other flow rate.
To return to the original calibration value, set the defaults. (See page 30"Data Deletion.")
FC-M1
Collection device
Buffer tank
MP-
∑
N
Ⅱ
Series