- 12 -
Installation and Piping Methods
Install the pump body on a flat location. When choosing an installation site, avoid humid locations or
locations that are splashed with water, locations near fire or heat generating sources, and extremely
dust locations.
The pump body can also be mounted on a tripod. Insert the tripod screws into the tripod mounting
hole on the base of the pump body. If a tripod with a mounting base over 40 mm is used (i.e. one that
will result in a mounting footprint of 20 mm or more wide from the tripod screw at its center), the
pump body sometimes cannot be installed stably.
Make sure that the filter element is attached to the suction holder. Also, replace the filter element
when it becomes particularly dirty. (See page 14 "How to Install the Suction Holder.")
5 mm dia. and 7 mm dia. tube can be inserted onto the suction port.
* When inserting piping, take care to prevent strong force from being applied to the suction holder.
Unnecessary force will damage the suction port.
Likewise, 5 mm dia. and 7 mm dia. tube can be inserted onto the exhaust port. However, pay
attention to the following points:
•
The piping itself sometimes acts as a load and prevents suction performance as indicated in the
specifications from being achieved.
•
The flow volume on the exhaust side cannot be controlled. Treat indicated flow volume values for
reference only.
During Ultra Low Flow Volume Measurement (supported only on MP-
Σ
30KN
Ⅱ
)
When measuring under a flow rate of 0.150 L/min or less, use the ultra low flow rate suction holder
that is provided.
Check the filter element in
the filter check window.
Regular
suction holder
The center O-ring
is transparent.
The center O-ring
is black.
An orifice is
contained inside.
Ultra low flow rate
suction holder