76
R: Loading Log Data
Computer Command Messages
<STX> ID R (1) <ETX>
BCC(H) BCC(L) <EOT>
(1) Number of loaded data records (4 characters) ······ "####"
(All data are loaded if "0000" or
exceeding the number saved)
Input example: 01R0100
FLD-1 Response
<STX> ID R Result
code (1) <CR><LF>
(2)
<ETX> BCC(H) BCC(L)
<EOT>
(1) Header information (Data separated by commas.)
x
Current date (10 characters) ··················· "20##/##/##"
x
Current time (8 characters) ···················· "##:##:##"
x
Measuring device model (5 characters) ···· "FLD-1"
x
Station ID (6 characters) ························ "######"
x
RS-485 ID (2 characters) ······················· "##"
x
Averaging period (2 characters) ·············· "##"
x
Number of data records (4 characters) ····· "####"
(2) Measurement data (Data separated by commas. New line for multiple log events.)
x
Date saved (10 characters) ···················· "20##/##/##"
x
Time saved (8 characters) ······················ "##:##:##"
x
Dust concentration (5 characters) ············ "#.###" or "##.##" or "###.#"
x
Flowrate (3 characters) ························· "#.#"
x
Atmospheric pressure (6 characters) ········ "####.#"
x
Heater temperature (2 characters) ··········· "##"
x
External air temperature (5 characters) ····· "+##.#" or "-##.#"
x
Relative humidity (2 characters) ·············· "##"
x
Wind speed (4 characters) ····················· "##.#"
x
Wind direction (3 characters) ·················· "###"
x
Alarm (16 characters) ···························· "0000000#########"*
* See the figure below for corresponding alarms. (0: No 1: Yes)
Response example:
01R02014/05/01,17:30:00,FLD-1,000101,01,60,0100
2014/05/01,15:00:00,0.021,1.7,1013.2,50,+24.0,45,02.5,170,0000000000000000