94
Installing the Dust Unit
(1) Place the dust unit into the main FLD-1 unit
and partially fasten it with the 4 screws
removed previously.
Note: Use a Phillips screwdriver (No. 2).
Note:If the previously removed screws are lost,
provide the following screws and
washers.
x
Four M3
×
8 mm Philips screws
x
Four M3 flat washers
x
Four M3 spring washers
(2) Rotate and slide the connector pipe down into
the valve until the line mark on the connector
pipe reaches the edge of the pipe inlet on the
valve.
(3) Finish tightening the screws that were partially
tightened above.
(4) Connect the connector on the end of the cable
from the dust unit into the connector on the
main FLD-1 unit.
Note: After connecting the connector, tighten
the front and back fastening screws.
Note: If the back screw is difficult to turn, try
using the fingers to turn the screw with
the connector positioned between your
two hands.
Note: Push any cable slack in toward the back of the unit.
Line mark on
connector pipe
Connector
Fastening
screws