62
FL4, НЕИСПР FL5 is deactivated. If in this case no echogram or symbol
report is recorded, СБОЙ FL16 or СБОЙ FL16 will be displayed when the
unit is initialised. If faults НЕИСПР FL4, НЕИСПР FL5, НЕИСПРFLES,
СБОЙ FL16, СБОЙ FL17 are displayed, the device must be sent for repair.
Supplement 13. List of possible malfunctions and methods of
rectification
Description of defects,
external manifestations
and additional signs
Probable cause
Rectification method
1.
_
No indication on the
display
The battery cell is
discharged
Charge the battery (see
Supplement 4)
2. Strong backwardness of
the date and time from the
current
Very deep discharge
of the
battery cell when the
tool has
been stored in a
storehouse
without charging the
battery
Charge the battery cell
(see
Supplement 4), transfer
information to
computer,
initial the tool, set the
current date and time
3. A slight time lag from the
current
Power outage of the
watch during battery
replacement
Clear the device
storage, set the current
date and time