49
Character
Setting Item, Function, Setting Range
Factory Default
Transmission output status in
Adjustment mode, Span sensitivity
correction mode
Last value HOLD
• Selects Transmission output status in Adjustment mode or Span
sensitivity correction mode.
• Available only when Transmission output (TA option) is ordered.
• Selection range
: Last value HOLD (Retains and outputs the last value
before adjustment or span sensitivity correction.)
:
Set value HOLD (Outputs the value set in [Set value HOLD].)
: Measured value (Outputs the value measured in
Adjustment mode or Span sensitivity correction mode.)
Set value HOLD
ORP transmission: 0 mV
MV transmission: 0.0%
• Sets the Set value HOLD.
• Available only when
(Set value HOLD) is selected in
[Transmission output status in Adjustment mode, Span sensitivity
correction mode].
• Setting range: ORP transmission: -2000 to 2000 mV
MV transmission: 0.0 to 100.0%
Backlight selection
All are backlit
• Selects the display to backlight.
•
: All are backlit.
: ORP Display
: Setting Display
: Action indicators
: ORP D Setting Display
: ORP D Action indicators
: Setting D Action indicators
ORP color
Red
• Selects a color for the ORP Display.
•
: Green
: Red
: Orange
: ORP color changes continuously.
The ORP Display color changes according to [ORP color
reference value] and [ORP color range] settings.
• When ORP is lower than [ORP color reference value] –
[ORP color range]: Orange
• When ORP is within [ORP color reference value]
[ORP
color range]: Green
• When ORP is higher than [ORP color reference value] +
[ORP color range]: Red
(Fig. 7.7-1)
Orange
Green
Red
: ORP color reference value
: ORP color range
Hys Hys
Hys