[About each mode and setting items]
(*1) In ORP Display mode or Cleansing Output mode, measurement starts, indicating
the item selected in [Backlight selection (p.49), ORP color (p.49), Bar graph
indication (p.50) and Setting Display indication (p.50)] in Special Function group.
If power is turned ON again, the last mode (ORP Display mode or Cleansing Output
mode) from when power was turned OFF will resume.
(*2) If
(Cleansing output) is selected in [EVT1 type to EVT4 type] in the EVT1
to EVT4 Action Groups, the unit will enter Manual Cleansing mode.
After cleansing action is complete, the unit automatically reverts to Cleansing
Output mode.
(*3) The unit cannot enter Adjustment mode or Span sensitivity correction mode in the
following cases:
•
(Lock 1)
,
( Lock 2) or
(Lock 3) is selected in [Set
value lock (p.47)].
• When
(Cleansing output) is selected in any of [EVT1 type to EVT4 type
(pp. 23, 29, 35, 41)], and cleansing action is performing using the ‘Cleansing time’
and ‘Restore time after cleansing’ settings.
(*4) Setting groups and items with dotted lines are indicated only when the EVT3 option
is ordered.
[Key Operation]
•
+
(3 sec): Press and hold the
key and
key (in that order) together
for 3 seconds. The unit will proceed to Manual Cleansing mode.
•
+
(3 sec): Press and hold the
key and
key (in that order) together
for 3 seconds. The unit will proceed to Adjustment mode.
•
+
(3 sec): Press and hold the
key and
key (in that order) together
for 3 seconds. The unit will proceed to Span sensitivity correction
mode.
•
+
(3 sec): Press and hold the
key and
key (in that order) together
for 3 seconds. The unit will proceed to Transmission output Zero
adjustment mode.
•
,
: Press the
or
key. The unit will proceed to the next setting
item, illustrated by an arrow.
•
,
: Press the
or
key until the desired setting mode appears.
• To revert to ORP Display mode or Cleansing Output mode, press and hold the
key for 3 seconds while in any mode