Mode -03- (Process Time Measurement)

The following parameters must be set in mode -03- so that the DT-5TXR/DT-5TS will display the correct readings:

Instructions for setting parameters in mode -03- are as

follows:
1. Press the PAR button; the display will indicate:

The secondary display A shows that the meter is in

mode -03-,  and  parameter  1  is  showing  that  the

primary display will count in hundredths of a second

(000.00).

2. Use  any  of  the  lower  buttons  to  toggle  between

the meter reading in seconds or reading in hours,

minutes, and seconds (see table above ).

3. Press the PAR button; the display will indicate:

Parameter  2  is  showing  on  the  primary  display

(1  [factory setting]).

4. The input signal edge can be triggered by the leading

edge (1) or trailing edge (0).  Press any of the lower

buttons to toggle between these two settings (see

table above).

5. Press the PAR button; the display will reflect:

“How to Operate” Teaching Feature

NOTE: This feature can only be used in mode -01-.
With the DT-5TXR/DT-5TS it is possible to monitor the speed

of a motor when the number of pulses that the sensor is

generating is unknown. The RPM can be programmed

into the meter if it is already known or discovered by

using a handheld tachometer (Shimpo models DT-105A

or DT-107A for contact measurement; DT-205LR or DT-207LR

for non-contact measurement). The meter will then use

this number to set parameters 1 and 2.

Set the teaching feature as follows:
1. Determine the output RPM.
2. Confirm that the meter is receiving a signal from the

sensor (the SIG indicator will be illuminated).

3. Press  the  TEA  button  (the  TEA  indicator  will  be

illuminated).

4. Use the NUMERIC buttons to enter the known RPM.
5. Press the SET button to save the setting.

The parameters for mode -03- are completely different for those of mode -01- and mode -02-.

The  primary  display  is  now  showing  parameter  3

(3600 sec [factory setting]).

6. Use the NUMERIC buttons to enter the amount of

time in seconds that the display needs to hold the

measured process.

7. Press the PAR button; the display will indicate:

Parameter 4 is showing on the primary display (10kHz

[factory setting]).

8. Refer  to  the  “Connection  of  Sensors”  chart  under

the  “Set-Up”  section  for  the  proper  filter  for  the

sensor being used.  Press any of the lower buttons

to scroll through the input filters (see table above)

until the desired reading is showing.

9. Press the SET key to save these parameter settings

and exit.

NOTE: Please see page 14 for an example of MODE -03-.

DISPLAY

MEANING

Hours, minutes, seconds

00:00:00

or

or

Hundreds of a second

0:00

0 = Trailing edge

1 = Leading edge

3

Auto zero time

0.1 – 3,600 seconds

_

3600.0

3600 seconds

4

Filter

10, 0.02 kHz

_

10

_

10 kHz

FACTORY SETTING

RANGE

SETTING

PARAMETER

2

d

n

o

c

e

s

 

0

0

1

/

1

1

0:00

_

1

 _

Leading edge

Input signal edge

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for DT-5TS

Page 1: ...Instruction Manual DT 5TXR DT 5TS DIGITAL PANEL MOUNT TACHOMETER www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...rtant Safety Precautions 3 Display and Keypad 4 Rear Panel and Terminals 5 Set Up 5 Power 5 Sensor Connection 5 Self Test 5 Panel Installation 6 Factory Settings 6 Parameter Settings 6 Function Settin...

Page 3: ...g while saving time Accepts modules DT 5TXR only which provide different types of outputs i e analog voltage analog current 4 20 mA relay closures set points BCD output and others for recording purpos...

Page 4: ...a r a p s t e S n o t t u B t e S 7 n o i t c e l e s e d o M n o t t u B e d o M 8 l a c e d r o f a e r A l a c e D t i n U 9 t n e m e r u s a e m f o s t i n u s y a l p s i D y a l p s i D y r a...

Page 5: ...nts etc 4 When satisfied that all LED s are operational press the SET button to exit SENSOR TYPE TERMINAL NUMBER FREQUENCY or RPMRANGE OPERATION TEMPERATURE FILTER M P R 0 0 0 5 0 C 0 6 B 1 E R 3 0 C...

Page 6: ...1 000 RPM 3 Display units 1000 1 000 4 Update time _ 1 0 _ 1 second 5 Display Hold after input signal ceases _ _006 0 6 seconds 6 Input filter _ 10 _ 10 kHz DEFAULT FACTORY SETTING PARAMETER DESCRIPT...

Page 7: ...ogrammed into the DT 5TXR DT 5TS In addition there are also function settings to further customize the DT 5TXR DT 5TS NOTE Because the flow rate measurement mode 04 is only used in the Japanese market...

Page 8: ...feet per minute and the belt that is being measured is moving at a rate of 55 5 f min then 5 5 5 would have to be entered and the negative decimal point button pressed once If the desired reading is R...

Page 9: ...Press the PAR button 8 Use any of the lower buttons to scroll through see table above and select an update time 9 Press the PAR button 10 Use the NUMERIC buttons to enter the time in seconds that the...

Page 10: ...e meter will then use this number to set parameters 1 and 2 Set the teaching feature as follows 1 Determine the output RPM 2 Confirm that the meter is receiving a signal from the sensor the SIG indica...

Page 11: ...e r e t s y H 2 _ __ _ 0 0 0 3 Secures the 1st high and 1st low setting 0 not secured 1 secured _ 0 _ Notsecured d n o c e s 9 9 0 t i m i l t e s y n a f o n o i t a v i t c a t u p t u O 4 _ __ _ 0...

Page 12: ...activate the MEM button by pushing it momentarily The lower left secondary display will reflect the maximum value of the main display and the lower right secondary display will reflect the minimum val...

Page 13: ...nd the operator wants to observe the changes in this case higher update times are selected or if the speed changes rapidly within certain limits this may bother the observer a lower update time is sel...

Page 14: ...DT 5TS has a MODE for this application and similar ones where it can measure the ON time of a signal that starts when the process begins until it ends see signal waveform diagram top right NOTE The di...

Page 15: ...20mA DOP CPTR Relay output HI GO LO 3 C type relays DOP TRTR NPN OC output H 1 H 2 GO L 1 L 2 and ZERO DOP RMTR Ratio of two input functions Lower Connector Options with Connectors C DOP FVC Analog ou...

Page 16: ...OP FVC module coupled with the DT 5TXR panel meter provides three popular voltage outputs and one current output for recording purposes and PLC applications The voltages are a 0 to 10 VDC b 1 to 5 VDC...

Page 17: ...ll be approximately 10 msec 1 indicates that the update time will correspond with the display update time already chosen see Setting Mode Parameters 4 7 Press the FUN button display will reflect The d...

Page 18: ...Module DT 5TXR BCD The model DT 5TXR BCD combination satisfies PCs and PLCs that require a BCD parallel signal The DOP BCD module can be configured with a negative logic output 0 HI and 1 LO or a posi...

Page 19: ...ncy 30 kHz 0 999999 with selectabledecimal point 0 008 1 digit DT 5TXR Display Range MODEL Mode r e t e M e t a R r e t e m o h c a T Elapsed Time Process Time 0 00 00 9 59 59 0 00 00 0 59 59 or or 0...

Page 20: ...NIFICATION LIMITATION OF DAMAGES Buyer agrees to indemnify and hold Shimpo Instruments harmless from and against all claims and damages imposed upon or incurred arising directly or indirectly from Buy...

Reviews: