11. Environment
Read Chapter 7 Before Use of
These Pages
11. Environment
11.1 Overview
Settings on the balance can be changed to compensate for the installation environment such as the
degree of vibration or air movement or for the purpose of weighing a solid, liquid or powder.
11.2 Stability and Response (Averaging)
It is possible to match the stability of the display and the degree of response with the requirements of
specific applications or the installation environment. One of the five modes can be selected. Note that
adjustments for stability and response conflict with each other, although the UW/UX series is designed
to meet both.
Auto mode:
Select
menu
item
.
The balance automatically performs optimum averaging dynamically while
observing the load data. This is the recommended setting and should be used
unless special circumstances exist.
Pouring mode
: Select
menu
item
.
This mode is only suitable for constant volume weighing of liquids or fluids.
This mode is very sensitive to wind and vibration.
Standard mode
: Select
menu
item
.
This mode is suitable for weighing in a normal environment. Averaging is
fixed and does not change dynamically as in the Auto mode.
Anti-vibration mode
: Select
menu
item .
Use this mode when the balance is used in a location where there are large
vibrations and the display fluctuates in the Auto mode.
Response is deteriorated at small mass amount changes.
Anti-wind mode
: Select
menu
item
.
Use this mode when the balance is used in a location where it is exposed to
airflow that causes the display to fluctuate in the Auto mode.
Response deteriorates further than the Anti-vibration mode, but weighing is
comparatively stabilized.
Note
If weighing cannot be performed efficiently even with the Anti-wind mode,
change the installation site of the balance or use the optional windbreak
(large).