46
13. Enhancing Productivity
Read Chapter 7 Before Use of
These Pages
13.1.1 Checkweighing (Comparator) Display Type 1
This is the best mode to determine pass or failure judgment based on the sample weight.
1
Select menu map item
.
2
Set the upper threshold value, which corre-
sponds to the upper triangle mark, with menu item
.
3
Set the lower threshold value, which corre-
sponds to the lower triangle mark, with menu item
.
13.1.2 Checkweighing (Comparator) Display Type 2
Use this mode for classification based on the sample weight. The display looks like a bar graph, but
also includes a checkweighing function.
1
Select menu map item
.
2
Set the upper threshold value, which corre-
sponds to the upper triangle mark, with menu item
.
3
Set the lower threshold value, which corre-
sponds to the lower triangle mark, with menu item
.
Upper threshold
<
Sample weight
HI
Lower threshold
≤
Sample weight
≤
Upper threshold GO
Sample weight
<
Lower threshold
LO
Displays in use
15
16
17
Note
Determination as follows:
Upper threshold
<
Sample weight
HI
Lower threshold
≤
Sample weight
≤
Upper threshold GO
Sample weight
<
Lower threshold
LO
Displays in use
18
19
20