When two minutes pass since mass display blinking start, "PLS.CAL" is displayed.
If "PLS.CAL" is displayed, please unload the thing on the pan and execute span adjustment.
*
Push key when you must keep measuring after “PLS.CAL” is displayed. Return to
blinking mass display.
If the span adjustment is not executed, the mass display blinking for two minutes and "PLS.CAL"
display are repeated.
PSC configuration procedures
1
Press and hold the
for approximately 3 seconds.
The calibration menu will open.
2
Press the
or
several times to display "PSC".
3
The stability mark will be displayed or turn off each time the
is pressed.
Stability mark
"PSC" setting
Off
OFF (Default setting)
Displayed
ON
4
Press the
several times or press and hold the
for 3 seconds
to return to mass measurement mode.
*
In the case of a verified balance as a legal measuring instrument,
is not displayed in the menu.
In the case of a verified balance as a legal measuring instrument, if one of the PSC start
conditions in (1) to (3) above is satisfied in the “ATY-R” series, the
symbol begins to blink.
Once 2 minutes elapses after the
symbol begins blinking, different operations will be carried
out as detailed below depending on if there is any load on the pan or not.
Case1) There is no load on the pan.
"PLS.CAL" is displayed. If "PLS.CAL" is displayed, press
and execute span adjustment.
Case 2) There is some load on the pan.
The
symbol and mass display will both blink. After 2 minutes have passed after the mass
display begins blinking, "PLS.CAL" will be displayed, so remove the sample from the pan and
press [CAL] to start span calibration.