144
V
GLP Output Function
13
FOR YOUR INFORMATION
3
Change the setting.
Pressing
alternately sets the ON and
OFF settings.
[SET]
4
Return to the weighing mode.
or
*
On setting the GLP output function to ON and outputting weight readings…
A long time is required to output one weight reading.
In addition, depending on conditions, data may not be printed correctly at the printer. See the setting
conditions below.
Handshake Settings
in the Communication
Settings (
^
page 129)
Rough Time Required for Output of One Weight Reading
Printer Only
PC Only
Both Printer and PC used
OFF
Approx. 10 sec.
Approx. 3 sec.
Approx. 10 sec.
SW (software)
Approx. 10 sec.
Approx. 3 sec.
Approx. 10 sec.
HW (hardware)
Approx. 10 sec.
Correct printing
is not possible
Approx. 10 sec.
TIM (timer)
Approx. 35 sec.
Approx. 35 sec.
Approx. 60 sec.
*
On setting the GLP output function to ON and using the command code "D01 (continuous
output)"…
No items other than weight readings are output.
! !S!E!T! !
!G!L!P.O!U!T
When ON is set the stability mark is lit.
Summary of Contents for ATX-R Series
Page 3: ......
Page 159: ...158 MEMO...