VACUUM LEAK DETECTOR VLR ..
2020-06-29
4
cal and audible alarm signal is triggered. If solenoid valves are connected, they close and the
pump stops.
If in place of the liquid stop valve a leak probe is mounted in the suction line in connection with
solenoid valves, the alarm signal is triggered when leak liquid reaches the leak probe. This
closes the solenoid valves and the pump stops.
3.4. Switching values of the leak detector in mbar
NOTE: The leak detector with the lowest alarm pressure for the respective application should be
used whenever possible (lower component wear)
Type
alarm ON
Pump OFF
Use on
Group:
VLR 230
> 230
< 360
P/Q
VLR 330, VLR 330/E
> 330
< 450
P/Q
VLR 410, VLR 410/E
> 410
< 540
P/Q
VLR 500, VLR 500/E
> 500
< 630
P/Q
VLR 570, VLR 570/E
> 570
< 700
P/Q
VLR .. - .. (/E)
Special switching values agreed between SGB and customer
The measured switching value for "alarm OFF" has to be at least 5 mbar smaller than the
measured switching value for "pump OFF".
The measured switching value for "pump ON" has to be at least 15 mbar bigger than the meas-
ured switching value for "alarm ON".
3.5. Description of the display and operating elements
3.5.1 States of the display elements (signal lamps)
Signal lamp
O
pe
ra
ti
ng
s
tat
e
S
tart
-up
S
tar
t-
up
, a
larm
a
c
k
no
wl
e
dg
es
A
lar
m,
v
ac
uu
m
be
low th
e
"al
arm
O
N"
l
ev
e
l
A
lar
m,
as
l
eft
-
ha
nd
c
ol
u
mn
,
ac
k
no
wl
e
dg
e
d
A
lar
m p
ro
be
A
lar
m p
ro
be
,
ac
k
no
wl
e
dg
e
d
A
lar
m s
ol
e
no
id
v
al
v
e
A
lar
m s
ol
en
o
id
v
al
v
e,
ac
k
no
w
l-
ed
ge
d
Dev
ic
e m
al
fun
c-
ti
on
OPERA-
TION: green
ON
ON
ON
ON
ON
ON
ON
ON
ON
ON
ALARM:
red
OFF
BLINK
ING
BLINK
ING
ON
BLINK
ING
OFF
OFF
ON
BLINK
ING
ON
3
ALARM 2
4
:
red
OFF
BLINK
ING
BLINK
ING
OFF
OFF
ON
BLINK
ING
ON
ON
OFF
Description:
Start-up: If the alarm is acknowledged as the pump is being put into operation, an optical dis-
tinction is not made, the audible signal is on or off depending on the button position. When the
switching value "alarm OFF" has been exceeded, the audible signal is always off.
3
The button "mute" is without function
4
Only applies to VLR ../E
Summary of Contents for VLR 230
Page 19: ...I I II II III III IV IV 21 21 21 21 20 20 20 20 27 11 2002 P 060 000...
Page 27: ...21 02 2017 SL 853 600 2 6 4 5 3 11 12 1 AC DC 105 106 102 71 01 69 24 1 24 3 59 09 76 17 60 Rp...
Page 29: ...22 84 85 85 22 52 98 98 0 Kl 1 0 17 12 2002 P 115 392 21 21 20 20 90 90...
Page 39: ...28 06 2005 D i hrbild Abmes 200 3 0 0 2 6 0 220 T 140 8 Dimension Drilling...