Figure 8 – Dimensions
2.2 Mounting
2.2.1 Wall
Panel
Dimensional drawings are given to aid installation and to determine the position of
mounting holes. See Figure 8.
a.
Position the analyser panel on a wall at eye level and as close as possible to the
sample source.
b.
Secure with four (4) each x 6mm bolts, leveling the analyser before securing.
c.
Use a spirit-level to level the instrument. This ensures that the sample liquid will
flow correctly through the instrument.
2.3
Hydraulic Connections (Wall Panel Only)
2.3.1 Procedure
a.
Connect a length of 8mm (5/16”) drain hose to the drain outlet on the analyser.
Route the hose to maintain a gravity fed drain (downward slope).
NOTE:
the drain connection is also the sample overflow connection and may be
recycled into the sample stream. Refer to Figure 1.
b.
Connect a length of 5/16" sample supply tubing to the source using the 1/4" NPT
connector. Connect the other end of the sample supply tubing to the analyser
inlet. The maximum desired pressure for the flow is 10 psig (0.6 bar). The sample
tubing is rated to 55 psig (3.8 bar).
c.
A sample tap assembly (figure 4, 18) attaching to the sample drain, is available as
an optional extra.