18
previous screen. At factory default, the software is programmed to automatically return to the main
screen after 60 minutes without user interface.
3.2.1 Main
Screen
The Main Screen is the default screen for viewing the residual value in digital format. The
residual value appears on the screen in four (4) x 1-inch high digits and a decimal; the
decimal place automatically shifts according to the residual value. The residual oxidant
and units (e.g. mg/l free Cl2) are displayed on this screen in smaller digits under the
residual value, and the current time and date are indicated at the top of the screen. The
software is programmed to automatically return to the main screen after 60 minutes
without user interface, unless the user is viewing the data in graphical format. From the
Main Screen, the user can initiate four selection options: Graphing, Calibration, Set Up, or
Diagnostics. If an error or alarm condition takes place, a warning message will appear
across the lower section of the screen, just above the user interface options.
Figure12 Main Screen
3.3 Set Up
The user set up screens are designed to allow for hardware and software adjustments to the
Cl1000 to configure the unit for site-specific operation. The following table summarises the user-
configurable options and the location of the screens for these adjustments:
Screen Name
Screen Function
Notes
Set Up
Lists adjustable Set Up Parameters
There are a total of 8 adjustable
parameters on the Set Up Screen (note
continuation arrows for additional options)
Buffer
In the manual buffer adjustment mode,
this screen allows the user to adjust
the buffer feed rate.
No buffer used on Cl1000B
Therefore not used in Cl1000B
Buffer Setpoint
Not used for Cl1000B
Setpoint Enter
Not used for Cl1000B
Screen
LCD settings, including screen
brightness, contrast and screen saver
are adjusted under this selection
Bright
Allows user to adjust screen brightness
Contrast
Allows user to adjust screen contrast
Lamp
The lamp saver and timeout are
adjusted under this screen
Lamp Selection
This screen allows the user to
configure the lamp to be always on, or
to go to low power after 10 minutes, 5
minutes, or 1 minute of inactivity. This
allows the user to adjust the life of the
lamp per operational requirements.
If the lamp is configured to go to low power
after a time period, any button press will
return the screen to the normal brightness.