19
Screen Name
Screen Function
Notes
Screen Timeout
The software for the Cl1000 is
designed to return to the digital display
on the main screen after a period of
inactivity when the user is in any of the
Calibration, Set Up, or Diagnostic
screens. This screen allows the user to
set this time period at 60 minutes, or to
turn the screen timeout off.
If amperometric titration is being used to
confirm calibration of the instrument, the
screen timeout should be set at 60 minutes
to allow time for the calibration results.
Date Settings
The current date and time can be
adjusted under this screen, along with
the date format
Time
The current time is adjusted from this
screen. The current time is indicated
on the top line of every screen.
Date
The current date is adjusted from this
screen. The current date is indicated
on the top line of every screen.
Date Format
The current date, which appears on
the top line of every screen, can be
configured to appear in
month/day/year, day/month/year, or
year/month/day format.
Outputs The
user
configures the two mA output
signals and the alarm relays under this
selection
mA Output A
The function (or range) and high and
low set points for mA Output Signal A
are configured under this selection
1. Each mA output signal is independently
configurable.
2. The mA outputs and relays are
configured under the Output Set Up
screens. These outputs can be tested
under the Diagnostics screen.
mA Output A
Function
The value of mA Output Signal A can
be configured for 4-20 mA, 0-20 mA, or
0-10 mA from this screen
mA Output A
Low
Setting
The residual value for the low mA set
point for Output A is configured from
this screen.
mA Output A
High
Setting
The residual value for the high mA
setpoint for Output A is configured
from this screen
mA Output B
The function (or range) and high and
low set points for mA Output Signal B
are configured under this selection