OPERATION
No. Module Name
Data type Function
Comment
18
TIME_CONTROL
Byte 1
Bit definition:
0.
Set time
The time is applied when changing the bit from
0 to 1.
Byte 2
Day
1-31
Byte 3
Month
1-12
Byte 4
Year
Year starting from 2000. 0 cor-
responds to the year 2000.
16 corresponds to 2016. Input starting from
year 2000 possible.
0-105
Byte 5
Hour
MBE Change must be performed after setting
the time to reset the timer.
The old time setting is retained if an invalid date
is transmitted.
0-24
Byte 6
Minute
0-59
19
CALIBRATION_
CONTROL
Byte 1
Bit definition:
The calibration starts with the change of Bit 0
from 0 to 1. A valid value for the calibration
strokes and calibration speed must be present.
The calibration value is saved with the change
of Bit 2 from 0 to 1. A value greater than 0
must exist in Float 4 and the calibration must be
completed for this.
Using Bit 3, the calibration can be deactivated
for a calibrated pump by setting the bit to 1. The
control is then performed with specification of a
stroke frequency.
0.
Start Calibration
1.
Cancel Calibration
2.
Save Calibration Value
3.
Calibration ON/OFF
Word 2
Calibration strokes
1-9999 strokes
Byte 3
Calibration speed
1-100%
Float 4
Calibration result
0.000 - max. value in l
20
FLASH_CONTROL
Byte 1
Bit definition:
0.
Signals ON/OFF
1.
Operating data ON/OFF
2-3
Write period
00
1 minute
01
5 minutes
10
10 minutes
11
30 minutes
21
DIACHANGE_
CONTROL
Byte 1
Bit definition:
The change is started with the change of Bit 0
from 0 to 1. The change is completed with the
change of Bit 1 from 0 to 1.
0.
Start of the diaphragm ch-
ange
1.
Change completed