Step by Step: Data Collection with LISST-Infinite Monitor , Master
Controller Mode
Step
Action
Result
1
Start LISSTInfMonitor program. On the Windows
Start menu click on Programs. Then Sequoia
Scientific and then LISSTInfMonitor.
Program should start and
show four plots with the
words “No Data Available”.
2 From
Settings
menu of LISSTInfMonitor program
select
Program Settings
.
Confirm that the
Time Between Samples
is set to
the desired value. When the system is
configured with the submersible pump running
continuously the time between samples can be
set to as little as five minutes. However, when
the submersible pump is configured to run in
AUTO mode, where it is controlled to turn on and
off automatically, the minimum recommend time
between samples is 15 minutes. Short time
intervals could cause damage to the submersible
pump.
Settings window will be
displayed.
Time between Samples set.
3 From
File
menu of LISSTInfMonitor program
select
Open Master Controller
. Data previously
acquired will be displayed on the four plots.
Commands will be sent to the instrument to
initiate sample. Text in the lower left of the
window will inform the user of the current status
of the data collection.
Data Collection started.
4
When the data collection is complete the
LISST-Infinite will transmit the data to the
Monitor PC where the LISSTInfMonitor
software will receive it and process the raw