86
LISST-100X User’s Guide
2) Make sure the current program settings are as desired and saved
as default settings (see SD).
BB
BioBlock Connection
Syntax: BB x or bb x, where x can be 0, 1, 10 or 11
Description: Configure the BioBlock Connection Setting and send test commands.
Input options are:
0 = BioBlock is not connected.
1 = BioBlock is connected.
10 = BioBlock is connected and issues a command to close the
BioBlock.
11 = BioBlock is connected and issues a command to open the
BioBlock.
Example: Input: BB 1
Output: BioBlock is now connected!
Cautions: None
SD
Store Current Settings as Default
Syntax: SD or sd
Description: Issue the SD command to verify the current settings. Then confirm (1
= yes, 0 = no) that these settings should be burned into the EEPROM
to be used with the AutoStart setting.
Example input:
SD
output: LISST-100X Current Status and Settings
Serial number = 1335
Firmware Version 1.997 Feb 16 2011 14:18:45
Operating Mode: Fixed Rate Mode
Start Condition: Delay Start with 3 minute delay
Stop Condition: Fixed Number Stop at 2 samples
Measurements per Average: 10
Sample Interval: 5
Battery Voltage is 7.28
Current Date/Time: Wednesday, August 17, 2011 12:24:14
Current Day of the Year: 229
Do you wish to set current settings to be defaults? (1=yes,0=no): [0] ?
1
Current Settings being saved to EEPROM...
Current Settings saved as defaults!!!
Cautions: If you change any of the settings after issuing the SD command you
must reissue
the SD command to save the updated settings.
Summary of Contents for LISST-100X
Page 2: ......
Page 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Page 4: ......
Page 5: ......
Page 7: ......
Page 9: ......
Page 10: ......
Page 12: ......
Page 19: ...LISST 100X User s Guide 7...
Page 20: ...8 LISST 100X User s Guide...
Page 61: ...LISST 100X User s Guide 49...
Page 65: ...LISST 100X User s Guide 53 Notes...
Page 71: ...LISST 100X User s Guide 59 Notes...
Page 92: ...80 LISST 100X User s Guide consumption...
Page 110: ...98 LISST 100X User s Guide...