Step 7: Wake up
instrument from
Sleep mode
The instrument is shipped in a low power Sleep mode to conserve
battery power. To wake up the instrument select
Wake Up LISST
from the
LISST
menu or press the
button. A dialog box will
appear counting down the maximum time required for the
instrument to wake up. The instrument sleeps for 30 seconds,
wakes up for 1 second to look for a wake up command and then
goes back to sleep. Therefore the wait could be as long as 32
seconds for the LISST-100X. Upon wake up the instrument status
will be displayed. If the cable connections are not correct or if the
battery is not installed a warning message will appear after the
time expires. The serial port settings can be checked by selecting
Serial Port Settings
from the
Communication
menu. The default
settings are COM1, 9600 baud, 8 data bits, No parity, 1 stop bit,
and no protocol.
Step 8: Open
Terminal window
The LISST-SOP has a terminal window that allows the user to
communicate directly with the instrument using a set of 2-letter
commands; see LISST-100X Command Summary on page 79
for details. To open the Terminal window select
Terminal
Window
from the
Communication
window or select the
button
form the tool bar. A window similar to the one shown below will
appear.
Summary of Contents for LISST-100X
Page 2: ......
Page 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Page 4: ......
Page 5: ......
Page 7: ......
Page 9: ......
Page 10: ......
Page 12: ......
Page 19: ...LISST 100X User s Guide 7...
Page 20: ...8 LISST 100X User s Guide...
Page 61: ...LISST 100X User s Guide 49...
Page 65: ...LISST 100X User s Guide 53 Notes...
Page 71: ...LISST 100X User s Guide 59 Notes...
Page 92: ...80 LISST 100X User s Guide consumption...
Page 110: ...98 LISST 100X User s Guide...