LISST-100X User’s Guide
99
Technical Specifications
Parameters
measured/derived
Particle Size distribution
Total Volume Concentration
Optical Transmission
Depth
Temperature
An external analog input (0-2.5V)
Two external digital I/O ports (5V logic)
Optical path length: 5 cm standard, 2.5 cm optional
Optical transmission: 12 bit resolution
Particle size range: Type B: 1.25 - 250 micron diameter, Type C: 2.5 –500 micron
Resolution: 32 size classes, log-spaced
Data storage memory: 1GB (~12 million measurements) standard.
Maximum sample speed: 1 size distributions per second (standard)
Temperature-sensor range: -10 to 45
o
C resolution: 0.01
o
C
Depth Sensor range: 0 to 300 m of H
2
O, resolution: 0.08 m of H
2
O
External power input: 12V nominal (9-15V max, 0.5 A min;
9-24V max, 0.5 A min
for units shipped after 1 March 2012
)
Dimensions: 13.33 cm (5.25")
x 80.7 cm (31.75") L
Weight in air: 12 kg (25 lb)
Weight in water: 4 kg (8.5 lb)
Depth rating: 300 m
Connector Endcap screws: Socket head cap screws #8-32 3/4” long
Endcap O-ring: Parker Part # 2-244
Summary of Contents for LISST-100X
Page 2: ......
Page 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Page 4: ......
Page 5: ......
Page 7: ......
Page 9: ......
Page 10: ......
Page 12: ......
Page 19: ...LISST 100X User s Guide 7...
Page 20: ...8 LISST 100X User s Guide...
Page 61: ...LISST 100X User s Guide 49...
Page 65: ...LISST 100X User s Guide 53 Notes...
Page 71: ...LISST 100X User s Guide 59 Notes...
Page 92: ...80 LISST 100X User s Guide consumption...
Page 110: ...98 LISST 100X User s Guide...