Part 1—The Diviner 2000 display unit and probe
Setup mode
Page 12
ACN 007 916 672
Sentek Pty Ltd
5.
Enter the last two digits of the year using the numeric keypad, e.g. 99
or 00.
6.
Press
ENTER
to save the new date setting.
The cursor moves to the air count setting to begin normalizing the
probe.
Normalizing the sensor
The sensor is normalized using a special normalization tube to set the air
and water counts. This is necessary because each sensor responds slightly
differently to air and water. By entering these counts into the equation as
described in Appendix A, it is possible to standardize all sensor readings,
i.e. readings taken by different sensors become comparable.
This results in meaningful data and flexibility in the use of probes.
Ø
To set the air count
1.
Use the arrow keys to move the cursor to the
Air Count
field.
The
Air count
function is highlighted.
2.
Place the probe inside the sealed normalization tube.
3.
Hold the probe and normalization tube in the air, clear of all solid
objects.
4.
Press
ENTER
.
Note:
The reading is taken over a 10-second interval and the probe must be
held in the air for this period.
5.
If 'failed' is displayed on the screen, repeat the air count, steps 3 and 4.
Ø
To set the water count
1.
Use the arrow keys to move the cursor to the
Water Count
field.
The
Water count
function is highlighted.
2.
Press
ENTER
at prompt.
3.
Insert the probe into the sealed normalization tube.
4.
Place the tube and probe into a 10 litre water-filled bucket.
Note:
Make sure that there is at least 10 cm (4 inches) above the top,
below the bottom and on each side of the probe.
5.
Wait until the measurement is complete.
6.
Repeat if an error message appears.
Summary of Contents for Diviner 2000
Page 6: ......