Appendix I
28 of 33
Service Manual for the LuMon™ System
8
A
PPENDIX
I
8.1
Spare parts and maintenance tools
For belts, belt connector, power cords, VESA Screw Set, Fuses for the LuMon™ Monitor and EIT Advanced Interface Set refer to
the Product Catalogue for the LuMon™ System
A complete list of services and spare parts is available from SenTec AG upon request (
N
OTE
Concerning equipment, accessories, disposable and parts refer to section ‘warnings and cautions’ of the User’s Guide for the
LuMon™ System
8.2
Interfaces - Pin assignment
The mains power supply of the LuMon™ Monitor is separated by two Means of Patient Protection (MOPPs) between the belt
connector socket (for the applied part, the belt) and the interface connectors.
The interface connectors (serial ports, USB ports, Ethernet port) of the LuMon™ Monitor are not separated from each other. If at a
time accessory equipment is connected to only one of the interface connectors no additional safety measures are necessary to
comply with the requirements of IEC 60601-1. If, however, accessory equipment is simultaneously connected to two or more of the
LuMon™ Monitor’s interface connectors additional safety measures may be required to be compliant with the requirements of IEC
60601-1.
In case of doubt consult qualified technicians
.
8.2.1
Belt connector socket (Redel)
Figure 8-1: Belt connector socket
Table 8-1: Belt connector port - signals
Pin(s)
Signal
Description
1
ETH RX+
Received Data
2
ETH RX-
Received Data
3
ETH TX+
Transmitted Data
4
ETH TX-
Transmitted Data
5
n.c.
-
6
GND
Signal ground
7
SHIELD
Shielding
8
+6V
+6V power supply
9
n.c.
-
10
+6V
+6V power supply
11
SYNC_IN
eit frame synchronization input
12
SYNC_OUT
eit frame synchronization output
13
-6V
-6V power supply
N
OTE
The belt connector port shall be used with SensorBeltConnector or LuMon™ Connector only.