Repairs of the LuMon™ Monitor
24 of 33
Service Manual for the LuMon™ System
5
R
EPAIRS OF THE
L
U
M
ON
™
M
ONITOR
Repairs described in this section do not require opening the LuMon™ Monitor’s cover and shall be carried out by qualified service
personnel only.
Repairs that require opening the cover of the LuMon™ Monitor are to be performed by
SenTec Authorized LuMon™ Technicians
only (i.e. trained by SenTec AG or accredited partners) and are described in detail within the Repair Manual
N
OTE
Prior to maintenance, service or repair all components of the LuMon™ System must be cleaned and disinfected. More details can
be found in section ‘cleaning and disinfection’ of the User’s Guide for the LuMon™ System
N
OTE
If equipment is send to a
SenTec Authorized LuMon™ Technician
, local SenTec representative or SenTec AG please refer to
section ‘instructions for repackaging and shipping’ of the User’s Guide
A complete ‘Certificate of Disinfection’
and, if applicable, the ‘Repair or Investigation Request Form’
must be provided
N
OTE
To ensure traceability – a regulatory requirement – SenTec AG needs to know the serial and LOT numbers as well as software
versions of all components of the LuMon™ System.
This means, if you replace parts of the LuMon™ System or if you perform a software update you must report the serial number of
the respective LuMon™ Monitor and the serial and LOT numbers of the replaced parts to your local distributor using the ‘Technical
Service Report Form’
For spare parts and test tools refer to section
Tools:
T10 Torx screwdriver and Slotted 4mm screwdriver
5.1
Prior to repair
Prior to repair, all components of the LuMon™ System must be cleaned and disinfected. Refer to section ‘cleaning and disinfection’
of the User’s Guide
Set the ON/OFF switch of the LuMon™ Monitor to OFF and disconnect all connections on the rear panel of the device.
5.2
Replacement – support foot
Follow the procedure described in section
1.
Remove the three screws with a T10 Torx screwdriver as shown on
2.
Slide out the damaged foot and slide in the replacement part. Only use the foot which is matching the LuMon™ Monitor
configuration:
LuMon™ ADULT
identifies Adults/Children configuration and
LuMon™ NEO
identifies the
Neonates/Infants configuration.
3.
Fasten the three screws with a T10 Torx screwdriver as shown on
Figure 5-1: LuMon™ Monitor's support foot
(A)
N
OTE
After repair verify that the replaced support foot matches the configuration of the LuMon™ Monitor.