Maintenance/Safety and Functionality Tests
Service Manual for the LuMon™ System
11 of 33
3.1.3
Labeling
Check presence/legibility of the label on the belt connectors.
The type label can be found on the MatchBox part of the belt connectors (
). Example labels are shown below.
A
C
D
E
J
H
G
B
F
L
M
I
A
C
D
E
F
G
H
I
K
B
A.
Product name
B.
Manufacturer symbol and address
C.
UDI data matrix code
D.
UDI human readable
E.
IP (ingress protection class) code
F.
Follow instruction for use
G.
Dispose of according to Council Directive
2012/19/EU symbol
H.
CE (‘Conformité Européene’) label and
notified body number
I.
SenTec logo
J.
Radio frequency symbol
K.
–
L.
Serial number symbol and number
M.
Reference symbol and number
A.
SenTec logo
B.
Product name
C.
Manufacturer symbol and address
D.
Reference symbol and number
E.
Serial number symbol and number
F.
CE (‘Conformité Européene’) label and
notified body number
G.
UDI data matrix code
H.
UDI human readable
I.
Follow instruction for use
J.
-
K.
Dispose of according to Council Directive
2012/19/EU symbol
The EU Authorized Representative (EC REP) label can be found on the cable close to the Redel connector
A
B
C
A.
Symbol for manufacturing date and
manufacturing date
B.
Symbol for EC Representative and the
address
C.
Medical device symbol
N
OTE
:
If the label is not present/legible then contact
SenTec Authorized LuMon™ Technicians.
3.1.4
Status indicator LEDs - SensorBeltConnector
Connect the SensorBeltConnector to the LuMon™ Monitor and switch on the monitor. During the power-up
verify, that the Status LEDs in the ControlBox and MatchBox show the following sequence: MatchBox
LED blue
ControlBox LED green
MatchBox LED red flashing
all LEDs off.
N
OTE
Any deviation from this sequence – e.g. ControlBox LED lighting up blue or flashing red or MatchBox LED
flashing blue – indicates that the SensorBeltConnector is defective and must be replaced (