System Configuration
111
Figure 6-6: Setting the Sampling Interval
Press the
+
and
–
buttons to toggle between
Normal
and
High
. Selecting High will double the
sample points if you need to have higher resolution of the reflected wave. Note that this also
increases data volume and may slow down data collection.
It is rare for the user to change this value, unless they have specific requirements.
The following table displays recommended temporal sampling interval for different frequencies:
Center Frequency
(MHz)
Recommended
Sampling Interval (ps)
12.5
6400
25
3200
50
1600
100
800
200
400
250
400
500
200
1000
100
6.1.5
Stacks
In noisy environments, one way to improve the signal to noise ratio is to collect more than one
trace at each position, average them, and then record the average trace; this is known as
Summary of Contents for pulseEKKO
Page 1: ...2005 00040 09 ...
Page 2: ...ii ...
Page 4: ...Warranty Confirmation iv ...
Page 10: ...x ...
Page 14: ...Introduction 2 ...
Page 72: ...Assembling Full Configurations 60 ...
Page 80: ...Getting Started 68 ...
Page 116: ...Tools and Preferences 104 ...
Page 180: ...Collecting Data 168 ...
Page 190: ...Map View 178 ...
Page 198: ...Managing Data 186 ...
Page 221: ...Care Maintenance 209 ...
Page 222: ...Care Maintenance 210 ...
Page 226: ...Technical Specifications 214 ...
Page 228: ...Appendix A 216 ...
Page 234: ...Appendix B 222 ...
Page 236: ...Appendix C 224 ...
Page 238: ...Appendix D 226 ...
Page 246: ...Appendix E 234 ...
Page 248: ...Appendix F 236 ...
Page 250: ...Appendix G 238 ...