![Sensors & Software pulseEKKO Product Manual Download Page 118](http://html.mh-extra.com/html/sensors-and-software/pulseekko/pulseekko_product-manual_1244152118.webp)
System Configuration
106
6.1 GPR Parameters
For each parameter described below, there is a Reset to Defaults button, which will reset that
parameter to the recommended value.
6.1.1
Frequency
Pressing Frequency will take the user to a screen where they can select the frequency of the
antennas/transducers they are using. Deciding which frequency to use for a survey depends on
the objectives of the survey. As frequency decreases, the depth of investigation generally
increases but spatial resolution decreases. Therefore, the ideal survey will be one that uses the
highest frequency that adequately penetrates to the target depth.
The following table offers a guide to frequency selection. The values are based on practical
experience and should be used as an approximation only.
Center Frequency (MHz)
Depth (m)
1000
0.5
500
1.0
250
2.0
100
5.0
50
10
25
30
12.5
50
The list of possible frequencies is shown in Figure 6-2.
Figure 6-2: Selecting the correct frequency
Summary of Contents for pulseEKKO
Page 1: ...2005 00040 09 ...
Page 2: ...ii ...
Page 4: ...Warranty Confirmation iv ...
Page 10: ...x ...
Page 14: ...Introduction 2 ...
Page 72: ...Assembling Full Configurations 60 ...
Page 80: ...Getting Started 68 ...
Page 116: ...Tools and Preferences 104 ...
Page 180: ...Collecting Data 168 ...
Page 190: ...Map View 178 ...
Page 198: ...Managing Data 186 ...
Page 221: ...Care Maintenance 209 ...
Page 222: ...Care Maintenance 210 ...
Page 226: ...Technical Specifications 214 ...
Page 228: ...Appendix A 216 ...
Page 234: ...Appendix B 222 ...
Page 236: ...Appendix C 224 ...
Page 238: ...Appendix D 226 ...
Page 246: ...Appendix E 234 ...
Page 248: ...Appendix F 236 ...
Page 250: ...Appendix G 238 ...