3-System Setup Parameters
pulseEKKO PRO
50
3.4
Survey Parameters
Selecting the Survey Parameters menu option from the System Setup main menu (See Figure 3-1 on page
31), will display a screen similar to Figure 3-8. This section details each parameter listed in this menu.
Figure 3-8: The GPR Survey Parameters menu. Use the Up and Down buttons to move to the desired parameter and edit the
value by pressing the A and B buttons.
3.4.1 Start Position
This is the starting position for the first trace. Normally, most lines start at position zero (0.0) and the
position of each trace increments by the Antenna Step Size (Section 3.2.5: P38). The Start Position value
can be changed to a non-zero and even a negative value if necessary.
It is assumed that the units for all position and distance quantities are the units specified under Positional
Units (Section 3.4.3: P50).
For a CMP/WARR measurement, the start position is the initial antenna separation.
3.4.2 Antenna Step Size
The Antenna Step Size can be changed in this menu as well as the menu under System Setup - GPR
Parameters (Section 3.2.5: P38).
3.4.3 Positional Units
Two distance units are available: metres or feet. This information is stored in the header file and is used in
plotting scales. All distance parameters in the other menus will be interpreted based on the setting of this
item.