ST-200 aQua
Operator’s Manual
65
If your required model is present in this screen then select the model by
pressing appropriate number, if not then press Y button to see more
models
1. K Na iCa Li pH *
2. K Na iCa pH * Cl
3. K Na iCa Li
* Cl
4. K Na iCa Li pH Cl
N
→
Exit
By selecting the parameters as per requirement & arrange the
electrode blocks as per parameters.
Note:
In 5
th
Model Li parameter depends on Na and iCa parameters,
If Na or iCa or both Na & iCa not calibrated, Li won’t be analyzed.
E) Select iCa Unit:
Depending up on user requirement can select iCa measuring units to
mg/dL or mmol/L. By default it is in mmol/L.
Press: MENU
MENU
1.DIAGNOSTICS
2.USER MENU
3.SETUP MENU
4.PARAMETERS
5.SELECT iCa UNIT
Press 5
↓
iCa UNITS
UNIT : mmol/L
Y → mg/dL
N→ mmol/L