OPERATING THE ANALYZER
3. Operating the Analyzer
A. Calibration
The instrument has to be calibrated when it is switched ON or it
automatically calibrates on set calibrated frequency after four hours or
eight hours as per set frequency.
Following is the calibration procedure:
A. Calibration
Switch ON the instrument
↓
INITIALIZING…
↓
↓
ACQUIRING RANGES
Please Wait…
↓
SENSACORE
System Uncalibrated
Please Calibrate….
Na* K* iCa* Li* pH* Cl*
Press CALIBRATE key
↓
CALIBRATE
Y→ Continue
N→ Exit
Press Y
Displays PUMP CALIBRATION STEPS
↓
↓
PUMPING AIR
↓
SELF CHECK
Searching pack