ST-200 aQua
Operator’s Manual
23
Eg.
Na = 33.76 mV
K = 49.54 mV
iCa = 54.33 mV
Li = 48.52 mV
pH = -62.32 mV
Cl = 49.73 mV
↓
ANALYZE SAMPLE
Na = 140.9
mmol/L
K
= 4.02
mmol/L
iCa = 1.25
mmol/L
Li = 1.15
mmol/L
pH = 7.516
Cl = 100.2 mmol/L
Result Outside the measurement ranges as follows will not be reported
WHOLE HAPERIN BLOOD
Na+
: 20.0 - 250.0 mmol/L
K+
: 0.2 - 40.0 mmol/L
iCa++
:
0
- 10.0 mmol/L
Cl-
: 20.0 - 200.0 mmol/L
Li
:
0
- 5.0
mmol/L
pH
:
5.0 - 9.0
SERUM, PLASMA & CSF
:
Na+
: 20.0 - 250.0 mmol/L
K+
: 0.2 - 40.0 mmol/L
iCa++
: 0
- 10.0 mmol/L
Cl-
: 20.0 - 200.0 mmol/L
Li
:
0
- 5.0
mmol/’L
pH
: 5.0 - 9.0
If iCa unit is in mg/dL, then the measurement range is 0-40 mg/dL.
C. Daily Cleaner
Execute Daily Cleaner Solution every day mandatory to remove protein
deposition from the sensors. The daily cleaner is required and is essential
for trouble-free analyzer operation. Following is the procedure of daily
cleaner.