ST-200 aQua
Operator’s Manual
58
INTERCEPT:
Intercept is used if all the three control values read either higher or lower.
Figure 18
(i.e.) y = mx + c
Where m is the slope of the line and C is the y-intercept, which is the y-coordinate of
the location where line crosses the y axis.
For increasing/decreasing all the control values to desired level we have to select
+ or -10
variation for
Na, Cl & + or – 0.5
variation in
K, pH, iCa, Li.
In the above graph of
intercept, Na is reading all the three control levels higher approximately by 10. In this
case we have to subtract intercept as 10.00 and verify it by running all three controls.
Similarly, for all the parameters intercept can be added according to their requirement
and respectively intercept ranges.
Note:
- By default Slope factor value is
01.00
& Intercept is
00.00
SLOPE/INTERCEPT
SLOPE FACTOR:
[ 01.00 ]
INTERCEPT:
+ [ 00.00 ]
Y→ ‘+’
N→ ‘- ‘
Select
+ or – 10
variation for
Na, Cl & + or – 0.5
variation in
K, pH, iCa, Li
according to
the necessity for result.
Press N
Summary of Contents for ST-200 aQua
Page 8: ...ST 200 aQua Operator s Manual 7 ST 200 aQua Figure 1...
Page 13: ...ST 200 aQua Operator s Manual 12 Figure 7...
Page 29: ...ST 200 aQua Operator s Manual 28 E Sub Menus i Diagnostics...
Page 30: ...ST 200 aQua Operator s Manual 29 ii User Menu...
Page 31: ...ST 200 aQua Operator s Manual 30 iii Setup Menu...
Page 32: ...ST 200 aQua Operator s Manual 31 iv Parameters Note Star Indicates Blank...