ST-200 aQua
Operator’s Manual
44
↓
PUMPING CAL A
↓
PUMP CAL A TEST
Pump CAL = 1545
N→ Exit
Instrument performs pump calibration using Cal A & displays the pump steps.
VALVE TEST:
VALVE TEST is used to clean the flow path of valve, by using DI water & syringe
FLOW
1.REAGENT PRIMING
2.PUMP CALIBRATION
3.VALVE TEST
N→ Exit
Press 3
↓
VALVE TEST
1.CAL A FLOW
2.CAL B FLOW
3.CAL A/B In Flow
4.AIR FLOW
N→ Exit
↓
Press 1
CAL A FLOW ON
60
N→ STOP
Clean the CAL A flow path of valve before counter reaches ‘0’.
↓
Press N
CAL A FLOW OFF
N→ Exit
Press N
↓
VALVE TEST
1.CAL A FLOW
2.CAL B FLOW
3.CAL A/B In Flow
4.AIR FLOW
N→ Exit