ENAMEL RATER MANUAL: V.10
21
4.2
SETTING UP THE STATISTICAL OUTPUT
When SW11/1 and SW11/2 on the applications board (see
Appendix A Fig 1) are set according to the following table statistical
information is available when the Batch/Stats key is pressed.
SW11/1
SW11/2
AVAILABLE STATISTICAL OUTPUT
Off
Off
None
Off
On
Average followed by Range
On
Off
Average followed by Standard Deviation
On
On
Average followed by 3 x Standard Deviation
The statistical information is available after the reading from any
can within the batch has been saved.
When the statistical information has been examined, 3 choices are
available:
i.
Press the END key to add further cans to the batch.
ii.
Press the SAVE key to download the batch data to a
printer (assuming the printer and are correctly
configured, see Section 6).
iii.
Press the CLEAR key to erase the batch information from
the 's memory.
4.3
THE LEVEL INTERLOCK
When SW11/5 on the applications board (see Appendix A Fig 1) is
set off the operation of the changes as follows:
i
The can no longer detect when a can change
has occurred and will not auto start after the change.
Note that the level check indicator continues to operate
normally.
ii
Enamel rating with the level check indicator on is allowed
and commences when START is pressed.