12
ENAMEL RATER MANUAL: V.10
2.4
CONNECTING A NON-SENCON PROBE
If you want to connect another probe to your for
specialised tests such as end, or side-seam enamel rating this is
always possible using the 5 pin plug supplied, but the following
points should be noted:-
i.
The cable diameter must be in the range 3mm to 5mm to
ensure correct sealing.
ii.
Other probes are unlikely to support body contact and
level checking and so these functions will be disabled.
iii.
If (ii) above applies, then the automatic restart facility after
each can will be disabled.
Soldering of the cable should be carried out by an electrician as
follows:-
Pin
Function
1
Main probe (electrolyte connection).
2
Level probe (if not used, connect to 1).
3
Body contact check (if not used connect to
4).
4
End or can body contacts.
5
Not connected.
The probe can now be connected to the via the probe
socket (refer to
Fig.3
). Ensure the connector is aligned before
insertion and screw in the locking ring finger-tight to ensure a good
seal.