24
Cryo Holder retraction:
1.
Close column valves, reset holder, objective out, flu screen down, cryo shield of holder closed.
2.
Gently brace against the purple plate of the goniometer and pull holder back until it stops, then
turn holder clockwise so that the “hat” rotates from the 12 o’clock to the 5 o’clock position
3.
Pull holder straight out of the column.
4.
Transfer holder immediately to the transfer station.
5.
If vacuum spikes and column pressure crashes, contact an EMG staff member immediately.
6.
Turn turbo on to prepare for next insertion.