Dry Pumping Station User Guide
Do not overtighten
the valves! Finger-
tight only!
Normal state at the
beginning of the day:
V1 = closed
V2, V3, & V4 = open
Removing holder
from dry pumping station:
1.
Vacuum should be reading 10
-3
to 10
-5
Pa.
2.
Remove temperature control cord by pulling straight out from base of plug from the cryo holder
jacket. Do not twist or pull the wire!
3.
Close V4, then close V3. If there is a second holder attached, close V4 and then V3 on that holder.
4.
Close V2. V1 should be closed already, but if not then close V1 last.
5.
Unhook tygon tubing from cryo holder and re-connect to silver stopping plug (behind V1 & V2).
6.
Turn pumping station off. Wait for green “MDP status” and “System Status” lights to turn off.
7.
Using one hand to stabilize the V1/V2 module, gently apply force to pull out cryo holder.
a.
Transfer cryo holder from pumping station into cryo transfer station.
8.
Place black blanking plug into opening where cryo holder was.
Inserting holder into dry pumping station and warming it up:
1.
Make sure all valves on pumping station are closed. If there are 2 holders, make sure the second
holder has V4 and V3 closed (in that order).
2.
Remove blanking plug from pumping station
3.
Insert cold cryo holder. Invert holder to facilitate condensation removal.
4.
Connect the plastic tygon tubing to V4 on the cryo holder.
5.
Turn on pumping station. Wait for green “MDP status” and “System Status” light to turn on (~1
min).
6.
Open V2, then V1 and wait until vacuum stabilizes.
7.
Connect cryo holder to the temperature controller and run the “WARMUP CYCLE” (~20 min)
a.
Warm up on the Elsa holder takes ~ 2 hr. Because of the long time to warm up, only keep
V1 and V2 open (V3 and V4 of BOTH holders CLOSED) during the warmup cycle.