45
9.4 Start: start condition of the data acquisition
•
Manual: with the
key.
•
Trigger: with a channel, with a combination of analog and logic channels (see chapter 5)
•
Stand-by: after a delay or at a precise date and time
9.5 Automatic: automatic stop when the block is full
1. Template channels: channels to compare with the template; access to the setup of the template
2. After acquisition: action after the end of the print
−
Save file: save the data acquisition into a file if the Real Time save has not been validated (or is impossible)
−
Draw: plotting the data acquisition
−
Stop: no other action
−
Relaunch: waiting for the Start trigger
−
Change setup: loading a configuration and waiting for the Start trigger
«
Impossible
» shows up when there is not enough place for saving data: you should reduce the sampling rate
or the number of channels
See chapter
File mode
to know the restrictions of the recording on file.
9.5.1 Creation of the template
Make a data acquisition into memory.
Press
, then select
.
The device displays the current template. It is marked as 2 curves (a min. curve and a max. curve) around the channel
selected as a reference for the template.
The 2 curves are saved in internal ROM.
1. Channel: allows the selection of the reference channel
2. DX & DY: allow you to build the min. and max. curves
3. Begin & End: limits of the memory extent where the trigger test shall be made
4. Save Go/NoGo: as soon as the template is correct, you can save it into the internal ROM.
You cannot save a template on a
file
. To save a template on a file, save the block and the channel in order to
recreate the template for reuse.