![SEFRAM BK Precision DAS30 User Manual Download Page 10](http://html.mh-extra.com/html/sefram/bk-precision-das30/bk-precision-das30_user-manual_1233326010.webp)
10
Introduction
The DAS30, DAS50 and DAS60 programmable recorders measures and records voltage, current, temperature, etc. on up
to 6 channels. The recorders also include digital input capabilities using the 12 logic input channels. An optional feature
of the recorders is a port for measuring temperature with Pt100 or Pt1000 temperature sensors.
1.1 Recording Modes
Finally, the optional thermal printer adds the ability to print recorded signals (saved and in real-time).
1.1.1 Direct mode
Print measured signals to the thermal printer
Use
Print readings and/or charts. Useful for recording long duration signals.
Enables
Complex triggers of the drawing, action after the end of the drawing, simultaneous recording to
a file.
1.1.2 Memory mode
Quick acquisition into internal memory of the measured signals
Use
quick short data acquisition (transient)
Enables
complex triggers of the data acquisition, action after the end of the data acquisition, simulta-
neous record on file
1.1.3 Go No-Go mode
Quick acquisition into internal memory of the measured signals
Use
quick short data acquisition (transient) for capture of non repetitive events
Enables
complex triggers, particularly when outrunning a pre-recorded template from a channel, action
after the end of the data acquisition, simultaneous record on file
1.1.4 File mode
Quick acquisition into internal memory of the measured signals
Use
quick protracted data acquisition (only restricted by the size of the saving means)
Enables
complex triggers, action after the end of the data acquisition, acquisition of very big amounts
of data.
1.1.5 Power Analysis mode
single-phase network analysis
three-phase analysis, Aaron mode, Star and Delta. (two-wattmeter method) .
Data acquisition on internal memory.
Requires Printer option installed at time of order