88
© Sealevel Systems, Inc.
SeaI/O | SL9049 11/2021
The SeaI/O-
470’s A/D channels can also be configured to provide up to eight 12
-bit current loop inputs. Each
input has two terminals
–
one positive and one negative. The input current range is 0-20mA for interfacing
commonly used 4-20mA devices. The dipswitches inside the enclosure must be properly configured for each
current loop input.
D/A Wiring Connections
The SeaI/O-470 provides two 12-bit D/A output channels, configured for 0-10V.
0-5V mode requires different jumper settings inside the enclosure. Refer to the Hardware Configuration
section of this manual for instructions on configuring the D/A jumpers.
The SeaI/O-
470 module’s D/A output channels can be independently configured for 0
-5V or 0-10V. The D/A
outputs do not support negative voltages.