72
Model 4688-IR / 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
CDR1
AL1
AL2
FAULT
0
OFF
OFF
OFF
1
ON
OFF
OFF
2
OFF
ON
OFF
3
ON
ON
OFF
4
OFF
OFF
ON
5
ON
OFF
ON
6
OFF
ON
ON
7
ON
ON
ON
ALARM
CDRBx
AL1 ON
CDRB0=1
AL1 OFF
CDRB0=0
AL2 ON
CDRB1=1
AL2 OFF
CDRB1=0
FAULT ON
CDRB2=1
FAULT OFF
CDRB2=0
Application Note:
When implementing the Alarm Test command (10), be sure to include the
Alarm Reset command (11). For example, in a graphical user interface this
might mean creating a "test" and "reset" button for each alarm (i.e., pushing
alarm 1 "test" button executes CDR1B0=1, FSR=10).
Command (11): Reset Alarms
Command (11): Reset Alarms
Command (11): Reset Alarms
Command (11): Reset Alarms
Command (11): Reset Alarms
Turns off the alarm override mode and clears the specified latched alarms.
Any alarms currently forced on or off by command 10 will return to normal
program control. This command may also be used in normal operation to
manually reset latched alarm 1 or alarm 2, but only after alarm conditions
have subsided. This function does not affect the RFF bit in the STS register.
Register sequence:
1. CDR1=0-4 (see table below)
2. PSR=not used
3. FSR=11
Coil sequence (for resetting one alarm only
1. CDR1B0 (or CDR1B1) = 1 (see table below)
2. PSR=not used
3. FSR=11
CDR1
AL1
AL2
0
No Change No Change
1
Reset
No Change
2
No Change
Reset
3
Reset
Reset
ALARM
CDRBx
Alarm1 Reset CDRB0=1
Alarm2 Reset CDRB1=1
See application note in Command 10.
Command (12): Set alarm inhibit timer
Command (12): Set alarm inhibit timer
Command (12): Set alarm inhibit timer
Command (12): Set alarm inhibit timer
Command (12): Set alarm inhibit timer
Updates the alarm inhibit countdown timer. The timer value is written into the
IDR1 register in units of 1 minute. Does not use PSR register. A timer value is
an unsigned integer therefore its maximum value is 65535 (1092.25 hours).