69
Model 4688-IR / Model 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
MODBUS
®
READ-ONLY
Register Descriptions
The READ ONLY registers contain most of the status data typically required
from most application using MODBUS
®
serial communications.
GAS: Gas concentration
GAS: Gas concentration
GAS: Gas concentration
GAS: Gas concentration
GAS: Gas concentration
Provides gas concentration reading in %LEL or %V/V. Concentration units
are indicated in the VOL bit of the transmitter status register, and are changed
through the SR.CU parameter in the USER menu.
Type
Real
Range
Programmed negative limit (AF.NL) to 9999
Access
Read-only
STS
STS
STS
STS
STS: T
: T
: T
: T
: Transmitter status
ransmitter status
ransmitter status
ransmitter status
ransmitter status
Provides information on status of various transmitter parameters and
indicators. AL1, AL2, and ALF are set when the corresponding display
indicator is active, while RL1, RL2, and RLF are set when the corresponding
relay is energized. MNT, IHB, and LCK are set when the MAINT, INHIB, and
LOCK flags are active. VOL and DGC is set when gas concentration units
and temperature are in %V/V and ºC, respectively. RFF is set when a data
access function has failed.
Type
Unsigned Integer
Range
0 to 4095 (12 bits)
Data high
0 0 0 0 RFF LCK DGC VOL
Data low
IHB MNT ALF AL2 AL1 RLF RL2 RL1
Access
Read-only
Note that in some circumstances relays may be energized even if indicators
are not displayed on the LCD - when alarm relays are programmed as fail-
safe.
TMP
TMP
TMP
TMP
TMP: T
: T
: T
: T
: Temperature
emperature
emperature
emperature
emperature
Provides temperature reading in °F or °C. Concentration units are indicated in
the DGC bit of the transmitter status register, and are changed through the
SR.TU parameter in the USER menu.
Type
Real
Range
-50°C (-58°F) to +448°C (838°F)
Access
Read-only