61
Model 4688-IR / Model 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
by the instrument. To ensure span accuracy the instrument's gas
concentration units (USER menu: SR.CU parameter) may be changed to %v/
v prior to beginning the span procedure. The F.CAL parameter is then
changed during the span procedure to match the %v/v value indicated on the
calibration cylinder (see Step 6).
STEP 6 - Initiate gas flow at 500 cc/min and allow gas to flow for 5 minutes or
until transmitter concentration reading is stable. Once stable, go to the
"
FACT
" menu, select the F.PWD (password) parameter and enter "251".
Upon reading "SEt" in the display, touch the magnet to the ESC key. Then
touch the magnet to the Up arrow key repeatedly until observing F.CAL. Touch
the magnet to the Enter key and adjust the reading to match the known
concentration of span gas applied to the sensor, then touch the magnet to the
Enter key and verify the display shows "SEt" Touch the ESC key 3 times to
return to "RUN" mode. Verify that the reading in the display matches the span
value within the stated accuracy of the instrument. Remaining in the "FACT"
menu, select and enter the "F.CAL" parameter. Using the Up/Down arrow
keys, select the concentration value of the calibration gas being used. Select
"Enter" to store the value and exit.
Note:
If span calibration for Methane(SR.GN = 1) or Butane (SR.GN = 4) is
performed, the cross calibration factors from Methane and Butane must be
copied to the other standard gases by using the "UP/DOWN" arrow keys. To
select F.WCF, touch the magnet to the ENTER key (1 or 4 will be displayed).
Touch the magnet to the ENTER key again and SET will be displayed when
successful. Touch the magnet to the "ESC" key 3 times until the display
shows "RUN".
STEP 7 - Turn off the span gas and reconfigure the instrument for on-line
service (remove cal cap and test adapter, etc.).
STEP 8 - Once the gas concentration reading has subsided to 0, restore
thealarms (local and remote) to their proper function (remove the inhibits).
NOTE: Do not
NOTE: Do not
NOTE: Do not
NOTE: Do not
NOTE: Do not
enter any menus
enter any menus
enter any menus
enter any menus
enter any menus
other than
other than
other than
other than
other than
specified without
specified without
specified without
specified without
specified without
approval from an
approval from an
approval from an
approval from an
approval from an
authorized Scott
authorized Scott
authorized Scott
authorized Scott
authorized Scott
Instruments
Instruments
Instruments
Instruments
Instruments
Service technician.
Service technician.
Service technician.
Service technician.
Service technician.
Damage to the
Damage to the
Damage to the
Damage to the
Damage to the
unit is possible.
unit is possible.
unit is possible.
unit is possible.
unit is possible.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com