68
Model 4688-IR / 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
MODBUS
®
Registers & Coil
Addressing
Register Address
Description
Suggested Tag
40001
Function select register
FSR
40002
Parameter select register
PSR
40003-40006
Character data registers
CDR1 - CDR4
40007-40010
Integer data registers
IDR1 - IDR4
40011-40014
Long data registers
LDR1 - LDR2
40015-40018
Real data registers
RDR1 - RDR2
40019
Gas concentration (read only)
GAS
40021
Temperature (read only)
TMP
40023
Transmitter status (read only)
STS
40024
General fault status (read only)
GFS
40025
Sensor faults status (read only)
SFS
Coil address
Description
Suggested Tag
00353
STS, relay 1 (1 = energized)
RL1
00354
STS, relay 2 (1 = energized)
RL2
00355
STS, fault relay (1 = energized)
RLF
00356
STS, alarm indicator 1 (1 = active)
AL1
00357
STS, alarm indicator 2 (1 = active)
AL2
00358
STS, fault alarm indicator (1 = active)
ALF
00359
STS, maint indicator (1 = active)
MNT
00360
STS, inhibit indicator (1 = active)
IHB
00361
STS, concentration units (1 = %v/v)
VOL (0 = %LEL)
00362
STS, temperature units (1 = ºC)
DGC (0 =
o
F)
00363
STS, lock indicator (1 = active)
LCK
00364
STS, remote function failure (1 = fail)
RFF
Appendix 1
READ
ONLY
READ
ONLY