48
Model 4688-IR / 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
Displayed
As:
Description
Min
Max
Default
Value
MEM
Region
Address
Data
Type
[BITS]
Assigned
Parameter
#
SR.tU
Sensor Temperature Units
- Determines
the units of measure for the temperature
reported under SR.TP:
0=°F / 1=°C
0
1
0 (°F)
PMEM
40336
Unsigned
Byte
[8]
31
SR.ZM
Sensor Zero Month (Read Only)
- When
the unit is zero calibrated, it stores the month
under this parameter. Since the unit is zero
calibrated during manufacture, the month will
be set at that time. The value is reset to 1
when default parameters are restored.
1=January
2=February, etc.
1
12
12
(updated
during
mfg.)
PMEM
40313
Signed
Byte
[8]
32
SR.ZD
Sensor Zero Date (Read Only)
- When the
unit is zero calibrated, it stores the date
under this parameter. Since the unit is zero
calibrated during manufacture, the date will
be set at that time. The value is reset to 1
when default parameters are restored.
1
31
31
(updated
during
mfg.)
PMEM
40314
Signed
Byte
[8]
33
SR.ZY
Sensor Zero Year (Read Only)
- When the
unit is zero calibrated, it stores the last 2
digits of the year under this parameter.
Since the unit is zero calibrated during
manufacture, the year will be set at that time.
The value is reset to 97 when default
parameters are restored.
0
99
97
(updated
during
mfg.)
PMEM
40313
Signed
Byte
[8]
34
USER MENU - Sensor Parameters
(continued…)