Model 4688-IR / Model 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
2.
Disconnect the ribbon cable located on the top (CPU) board in the
transmitter enclosure. Grasp the board and pull outward.
3. Make power, loop, and relay wire connections (18 to 22 AWG wire is
recommended), then connect the 9 conductor (master-to-remote) cable
as shown in Figure 10. Replace the CPU board.
4. Unscrew the (remote display) transmitter cover and remove the display
board. Connect the 9 conductor (master-to-remote) cable as shown in
Figure 10. Replace the display board and screw the enclosure cover
on, ensuring a tight seal.
5. Turn the power switch to the ON position on the (master) transmitter and
screw the enclosure cover on, ensuring a tight seal.
Powering The Transmitter
When power is applied to the transmitter,
all segments and indicators on the
display will turn on for 2 seconds, then will turn off for 2 seconds. The
transmitter will subsequently enter a 30 second warm-up period
(countdown shown on the display).