Model 4688-IR / Model 4679-IR
Operation & Maintenance Manual
Manual #087-0007G, 10/2003
Displayed
As:
Description
Min
Max
Default
Value
MEM
Region
MODBUS
Register
Address
Data
Type
[BITS]
Assigned
Parameter
#
SC.Md
Serial Communication Mode -
This
parameter specifies which communication
protocol to use: 0=None
1=Modbus
TM
ASCII
2=Modbus
TM
RTU
0
2
2(Modbus
™ RTU)
PMEM
40290
[hi byte]
Unsigned
Byte
[8]
78
SC.OP
Serial Communication Options
- Primarily
used in ASCII mode, this parameter controls
whether the transmitter echoes every
character sent, and if it expands outbound
carriage returns to carriage line feed.
This parameter should be set to 0 for Modbus
communication.
0 = no echo, no CR expansion
1 = echo, no CR expansion
2 = no echo, expand CR to CR/LF
3 = echo, expand CR to CR/LF
0
3
0
PMEM
40290
[lo byte]
Unsigned
Byte
[8]
79
SC.P
Serial Communication Parity Checking
-
Specifies whether parity checking is
performed and generated.
0=None
1=Odd
2=Even
0
2
0
PMEM
41288
[hi byte]
Unsigned
Byte
[8]
80
SC.RS
Serial Communication RS-232/ 485
Selection
- Specifies either RS-232 or RS-
485 communication. When choosing RS-
232, verify the jumpers at JP1 on the power
supply board are cut and jumpered as shown
in the electrical connections. Since RS-485
is the default, no modifications to JP1 are
required.
0 = RS-232
1 = RS-485
0
1
1 (RS-485) PMEM
40287
[hi byte]
Unsigned
Byte
[8]
81
SC.Rt
Serial Communication Retries -
This
parameter is not currently used.
0
9999
0
PMEM
na
na
na
SC.Sb
Serial Communication Stop Bits
-Specifies
either 1 or 2 stop bits
0=1 bit
1=2 bits
0
1
0 (1 bit)
PMEM
40289
[hi byte]
Unsigned
Byte
[8]
83
SC.VF
Serial Communication Verify
- When set to
1, turns off LRC checking in Modbus™ ASCII
protocol.
0
1
0
PMEM
40291
[hi byte]
Unsigned
Byte
[8]
84
SC.LO
Serial Communication Listen Only
- Setting
this parameter to 1 forces the transmitter to
listen for and process Modbus™ queries
(including broadcast), but without generating
a response.
0=listen and respond
1=listen only
This parameter should normally be set to 0.
0
1
0
PMEM
40291
Unsigned
Byte
[8]
77
COMM MENU - Serial Communication