SME1202A
1.2 Technical Specifications
Parameter
Specifications
Resistance Test
Range
100k
Ω
– 10T
Ω
Accuracy
>10nA:±2%,
≤
10nA:±5%
Current Test
Test Ranges (1-6)
Range 1 : 100µA–1mA, Internal I/P Impedance 10k
Ω
Range 2 : 10µA–100µA, Internal I/P Impedance 10k
Ω
Range 3 : 1µA–10µA, Internal I/P Impedance 10k
Ω
Range 4 : 100nA–1µA, Internal I/P Impedance 10k
Ω
Range 5 : 10nA–100nA, Internal I/P Impedance 1M
Ω
Range 6 : 1nA–10nA, Internal I/P Impedance 1M
Ω
Accuracy
2%±3pA
Test Voltage
Range
1V-1000V
Accuracy
≥
10V: 1% ±1V, <10V: 10% ±0.1V
Current Limit
10mA
ON/OFF
Manual On/Off on front panel, or controlled by built-in
timer, or by remote control
Charge Time
0-999s programmable
Measurement delay
0-999s programmable
Measurement Speed
Fast
≤
30ms, Slow
≤
60ms
Comparator Function
4 bins : 3 bins for PASS, 1 bin for FAIL
Range Mode
Auto, Hold
Memory
Internal memory and external USB disk
Standard Interface
RS232C, USB Host, USB Device, Handler
General Specifications
Environmental Conditions
Operating Temp : 0°C - 40°C, RH :
≤
90%RH
Storage Temp : 5°C - 40°C, RH :
≤
85%RH
Input Power
198 – 242V AC, 50Hz±5%
Power Consumption
<50VA
Dimensions
W:235mm, H:104mm, D:360mm
Weight
3.6kg Approx.
Accessories
Standard : Two terminal test clip leads, CD containing
User Manual
Optional : Foot Switch
1.3 Unpacking
Inspect the shipping container for damage after unpacking it. It is not recommended to
power on the instrument in the case of a damage container.
If the contents in the container do not conform to the packing list, notify us or your dealer.